Read The Isr, Iscr, Or Isce; Load The Isce; Fault Queue - Fluke 5730A/03 Operator's Manual

Multifunction calibrator
Hide thumbs Also See for 5730A/03:
Table of Contents

Advertisement

ZERO CAL
ACXFER
SETTLED
REMOTE
WBND
SCALE
OFFSET
PLOCK
PSHFT
RLOCK
RCOMP
BOOST
EXSENS
EXGARD
OPER

Read the ISR, ISCR, or ISCE

To read the contents of the ISR, send the remote command, ISR?. In a similar fashion, to
read the contents of the ISCR, send ISCR?, and to read the contents of the ISCE, send
ISCE?. The Calibrator returns a decimal number representing bits 0 through 15. Each
time the ISCR is read, its contents are zeroed.

Load the ISCE

By resetting the bits in the ISCE, the associated bits in the ISCR can be masked
(disabled). For example, to cause an SRQ interrupt when an attached 5725A Amplifier
turns on, bit 3 (BOOST) in the ISCE register must be 1. (The ISCB bit must also be
enabled in the SRE.)

Fault Queue

When a command fault, execution fault, or device-dependent fault occurs, its fault code is
placed in the fault queue where it can be read by the FAULT? command. All fault codes
are defined in Appendix A of this manual. Another way to decode a fault code is to send
the command, EXPLAIN?, which returns a description of a fault code. Reading the first
fault with the FAULT? command removes that fault from the queue. A response of 0
means the fault queue is empty.
The fault queue contains up to 16 entries. If many faults occur, only the first 15 faults are
kept in the queue. A 16th entry in the queue is always a "fault queue overflow". fault, and
all later faults are discarded until the queue is at least partially read. The first faults are
kept, because if many faults occur before the user can acknowledge and read them, the
earliest faults are the most likely to point to the problem. The later faults are usually
repetitions or consequences of the original problem.
The OPER command is inhibited for outputs of 22 V or greater whenever there is a fault
in the fault queue. The OPER command remains inhibited until either the fault queue or
the ESR is cleared.
When 1, DC Zero Cal is necessary.
When 1, ac/dc transfer is active.
When 1, the output has stabilized to within specification.
When 1, the Calibrator is under remote control.
When 1, the wideband is active.
When 1, scaling is active.
When 1, an offset is active.
When 1, the Calibrator output is phase locked to an external source.
When 1, variable phase output is active.
When 1, the Calibrator output range is locked.
When 1, two-wired compensation is active when in resistance mode.
When 1, an auxiliary amplifier is active.
When 1, external sensing is selected.
When 1, external voltage guard is selected.
When 1, the Calibrator is operating, When 0, it is in standby.
Remote Commands and Syntax
Check the Calibrator Status
6
6-55