Set a Static Internet Address

The 5730A Calibrator comes from the factory with 169.254.001.001 in the static IP
address register.
If the Calibrator is to be used on a corporate LAN and DHCP is not used,
contact the network administrator for a static IP address to be used
exclusively by the Calibrator. DHCP must be disabled to set a static IP
address.
To change the Calibrator static IP Address:
1. Touch Setup Menu>Instrument Setup>Remote Port>Ethernet Setup to get to the
2. Touch IP Address.
3. Use the numeric keypad to enter the IP address and then push .
The IP address is stored in non-volatile memory, and does not change when
power is removed and reapplied to the Calibrator or when the Calibrator
receives an *RST command.

Configure the General Network Socket Port

In order to communicate with each other, a client computer and the 5730A Calibrator
must use the same socket port number. The default port is 3490. Typically, the default
port does not need to be changed. If the socket port must be changed, enter the Socket
Port number supplied by the network administrator.
To change the Socket Port number:
1. Touch Setup Menu>Instrument Setup>Remote Port>Ethernet Setup to get to the
2. Touch PORT.
3. Use the numeric keypad to enter a new port number.
4. Push . The port number must be between 1024 to 65535.
The Network Socket Port Number is stored in non-volatile memory.
5-8
Ethernet Setup menu.
If a mistake is made when the IP address is entered,  must be pushed.
Ethernet Setup menu.
If a mistake is made when the socket port is entered, push  and go back to step 2
to enter the port number again.