Read the UUT Error: Resistance Output

To read the error of the UUT in the resistance function:
1.
Set the necessary resistance output as described under "Set the Output."
2.
Use the output adjustment controls as necessary to achieve a reading on the display
(illustrated below) equal to the reading on the UUT. To adjust a higher-order digit,
push . As the reference value is approached, work progressively back towards
the least-significant digit on the 5730A Calibrator display as necessary by pushing
. It is only necessary to go one digit past the least significant digit of the UUT.
3/15/2013
Standby
Reading
1.9000206

Introduction to Offset, Scale, and Linearity Errors

The performance of a DMM or meter can be plotted as a graph of input stimulus and
meter reading. For a perfect meter, the graph of the input stimulus would exactly match
the graph of the reading.
The 5730A Calibrator measures and shows three types of UUT errors:
Offset Error
Scale Error
Linearity Error
10:52am
1.9000096
42.00000
1.0000
External
External
Sense
Guard
OFF
OFF
Introduction to Offset, Scale, and Linearity Errors
+/- 10.0 ppm (1 year) 213 days since calibration
Nominal Value 1.9 kΩ
dBm
kHz
Error
= +5.79ppm
2 Wire
Wideband
Comp
ON
OFF
Front-Panel Operation
Setup
Menu
hhp133.eps
4
4-35

Advertisement

Table of Contents
loading

This manual is also suitable for:

5730a5730a/05

Table of Contents