Section 5

Setting the Number of Samples to Average

Setting the Number of Samples to Average
The average of the measured values obtained based on the preset number of samples can be
output. This setting can be used when you want to ignore rapid changes in measured values.
Note, however, that if this setting is made, the response time for judgment and linear outputs
drops.
Selection for No. of
samples to average
1
2
4
8
16
32
64
128
256
512
1024
Moving to the FUN mode and AVE
1.
Set the mode switch to FUN.
2.
Use the LEFT and RIGHT keys to display
[AVE] on the main display.
Selecting the number of samples to average
3.
Press the UP or DOWN key.
The sub-display flashes.
4.
Use the UP and DOWN keys to select the
desired number of samples to average.
5.
Press the ENT key to confirm the setting.
This registers the setting.
ZJ-SD
52
User's Manual
In the case of the static electricity sensor head, the
Response time
number of samples to average has been set to 64
before shipment from the factory.
2 ms
3 ms
5 ms
9 ms
17 ms
33 ms
65 ms
129 ms
257 ms
513 ms
1025 ms
In general, when the number of samples to average is
multiplied by "n", the resolution increases by
n
.