Two-point Scaling
Measurement is performed at two points and offset values are set for those measured
values.
Both an overall offset can be set and the range can be changed.
Displayed value
(kV)
Allow at least 1 kV between the two points to specify.
Setting the first point
1.
Set the first point by following steps
of the one-point scaling procedure.
p.99
The span for two-point scaling is automatically
set based on the values entered for the two
points. Inverted display settings are ignored.
Setting the second point
2.
Place the sensing object in the position (second point) for which the display is to be
changed.
The sensing object must be set at a distance at
least 1% of the rated measuring range away
from the first point and also at a distance within
the measuring range.
Changing Display Scales
Actual measured value
(kV)
1.
11.
to
Section 6
ZJ-SD
103
User's Manual