Cross Hairs Adjustment; Image Orientation Check - Siemens LITHOSTAR Multiline Adjustment Instructions Manual

Hide thumbs Also See for LITHOSTAR Multiline:
Table of Contents

Advertisement

2 - 8

Cross hairs adjustment

The shock wave head is in the coupling position (not to be confused with the GEL posi-
tion).
Switch fluoro on and check the isocenter at -30° and -10°.
Fig. 10
If the cross hairs need to be adjusted, the I.I. collision protection and the grid cover ring
on the I.I. must be removed.
CAUTION
Bypass jumper X18, so that the safety circuit is activated.
Loosen the screws in the retainer ring far enough so that the cross hairs can be adjusted.
Switch fluoro on and move the centering cross hairs until the ball of the adjustment
device is displayed in the center of the marking in both projection planes (-30° and -10°).
Tighten the cross hairs mounting screws. Ensure that the retainer ring does not bend.
Insert the I.I. grid at an angle of 45° ( ± 10°) between the bolts of the retainer ring.
Reinstall the I.I. collision protection and tighten the grid cover ring on the I.I. from below
to secure the grid and to prevent twisting.

Image orientation check

Close one of the collimator pairs completely.
Open the other collimator pair completely.
Release fluoro.
Check whether the image orientation is vertical or horizontal on the monitor.
To adjust, loosen the mounting screws on the camera and turn the camera head.
Check the image orientation again with FL .
LITHOSTAR Multiline
Centering the X-ray tube / Image intensifier
Acceptable
ok
The grid is attached by means of this cover and must also be removed.
Register
RXL2-120.071.01
Rev. 01
05.97
Correction
needed
Page 8 of 10
TD SP 1
Medical Engineering