Image Quality; Problems During Calibration; Quadrant Difference On Patient Images; Quadrant Missing - Siemens MAMMOMAT Nova Opdima 1000 Service Instructions Manual

Table of Contents

Advertisement

9 - 8

Image Quality

Problems during calibration

Check that grid lines are not visible in images taken with a table without grid e.g. biopsy
unit. If visible, see CCD camera calibration and maintenance on Page 5 - 22.
Calibration not completed takes unexpectedly long time. This can be a defective camera
or a mechanical alignment problem. Try to calibrate without external collimator, activate
switch or use a collimator with larger hole e.g. Axilla collimator.
NOTICE

Quadrant difference on patient images

When doing images on calibration plexiglas (also called: homogeneous acrylic plastic =
PMMA = Poly-methyl-met-acrylate) plates there can be a slight difference between the
quadrants, which will not be visible in patient images. Make sure that quadrant difference
exists in patient images before changing the CCD camera.
Calibrate the CCD camera.
If it is not disappearing, please contact the headquarter support center, HSC 24/USC, for
investigation.

Quadrant missing

Try to calibrate the CCD camera.
Please contact the headquarter support center, HSC 24/USC, for investigation.

White line/dot

Calibrate the CCD camera, move acrylic plastic plates after each calibration exposure.
If it is not disappearing, please contact the headquarter support center, HSC 24/USC, for
investigation.
Opdimar
Service
Always use Opdimar collimator in normal use.
Register 5
SPB7-230.061.11
Rev. 02
Troubleshooting guide