Setting A Specimen; Viewfinder; Measurement - Konica Minolta CM-26dG Instruction Manual

Hide thumbs Also See for CM-26dG:
Table of Contents

Advertisement

Setting a Specimen

Refer to P.33 for preparations to perform before measurement according to the specimen being measured and
the application.
Switch the measurement area (CM-26dG/CM-26d).
1.
Replace the target mask (CM-26d).
2.
Set the instrument on top of the specimen.
3.
„ Viewfinder
Switching the measurement area allows users to check the measurement point of a specimen when aligning
the instrument and the target are necessary, such as when the measurement location is small.
How to Use
1
Set the instrument on top of the specimen.
2
Slide the viewfinder lever to open the viewfinder.
3
As the viewfinder opens, a white LED will turn ON, illuminating the specimen within the range
of measurement.
Memo
When using the CM-26dG, the range of measurement will be illuminated according to the set measurement area
(MAV/SAV) when the measurement button is pressed.
4
Look into the viewfinder and adjust the position of the specimen.
5
Close the viewfinder. (The white LED will turn OFF.)
Notes
If the reflectance of the specimen to be measured is low, the measurement range may not be clearly
visible even when illuminated by a white LED. Also, if the specimen has a mirror surface or a surface
similar to a mirror surface, the illumination light may not be clearly visible. In such cases, use the optional
stapler type target mask (CM-A268).

Viewfinder

Viewfinder lever
2