… SMC Threshold

Sets the threshold to be used for the sample SMC function.
Memo • The SMC threshold is set to "0.4" when the instrument is shipped from the factory.
• For the threshold value, enter a value equivalent to the desired σΔE*ab for the average measurement data group
(average data entries set on P.91).
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "SMC
Threshold", and then press the [Confirmation] key.
2
▲ and ▼ are displayed above and below the number
to be set.
Use [▲] or [▼] to specify a value.
Settings
{ 0.01 to 9.99
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Measurement setup> screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Measurement setup> screen.
Start the procedure from the <Measurement cond.> - <Measurement setup> screen.