„ User calibration
You can perform calibration by using your own reference plate and calibration data instead of white calibration.
The calibration data for user calibration can be specified by connecting the instrument to a computer and using
the optional Color Management Software "SpectraMagic NX". Select whether to use the user calibration data for
measurement with <Calibration setup> - <User calibration> (refer to P.128) on the instrument.
Operating Procedure
1
Perform user calibration instead of white
calibration.
Notes
Before user calibration can be set to ON,
user calibration data must be prepared and
written to the instrument memory from a
computer.
2
Set the instrument on the user tile so that the
measuring port is over the tile.
3
Press the measurement button.
User calibration will be performed.
When user calibration is completed, the screen
returns to the one displayed before the
<Calibration> screen.
39
Start the procedure from the measurement screen.
Please set the
instrument on the user
tile and press the
measurement button.
User ID: xxxxxxx
CALIBRATION
Calibrating to user std