Flow of Measurement
Optional settings
↓
Language selection (
↓
↓
Measurement area selection (
↓
Zero calibration (
↓
↓
Gloss calibration (* CM-26dG only)
↓
* When color difference is
measured
↓
Color difference target settings
P.41)
(
(Procedure)
(1) Open the <Target>
screen.
(2) Select the target number.
(3) Measure the color
difference target.
↓
Pass/fail judgment criteria settings
(
P.69)
↓
↓
29
Basic procedure
Turning the Power ON (P.32)
P.103)
* As necessary (after initialization, etc.)
Measurement cond. setting (
* When required depending on the
P.33)
measurement conditions
* Only when required (when changing
P.35)
the measurement conditions, etc.)
White calibration (
P.37)
* When color difference
P.41)
Measurement (
(Procedure)
(1) Open the <Sample> screen.
(2) Measure the sample.
End
P.92)
↓
is not measured
↓
Simple measurement
Turning the Power ON
(P.32)
↓
Skip the initial
calibration screen
↓
When "Change to
simple mode?" appears,
select "Yes"
↓
Measurement
(Simple mode) (
P.48)
↓
End
* Configure measurement/
observation conditions
and perform calibration
as necessary.