… Tolerance Settings

Specify the tolerance used for pass/fail judgment of measured for each target.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Edit
tolerance", and then press the [Confirmation] key.
The tolerance list screen is displayed.
2
Use [▲] or [▼] to move the cursor to the item to set,
and then press the [Confirmation] key.
The <Edit tolerance> screen is displayed.
3
Use [▲]/[▼] and [◀]/[▶] to move the cursor to the
item to set, and then press the [Confirmation] key to
change the setting.
• When the selected item is not currently checked,
pressing the [Confirmation] key causes the item to be
checked, after which the setting value can be changed.
• Press [◀] or [▶] to move the cursor to the setting value
area. Pressing the [Confirmation] key causes the
cursor to appear on the value. Press [▲] or [▼] to
change the value.
Move the cursor between the digits of the value by
pressing [◀] or [▶].
Settings
{ x / y
{ Equation / MI
{ Other than the above : -20.00 to 20.00
• Press the [Confirmation] key to confirm every time the
setting of an item is changed.
4
After all settings have been configured, use [▲] or
[▼] to move the cursor to "OK", and then press the
[Confirmation] key.
The setting is confirmed and the screen returns to the
<Pass/Fail> screen.
Pressing the [ESC] while configuring the settings will prevent
the settings on the currently opened screen from being
applied and revert all settings to the previous condition.
5
Press [ESC] to return to the <Pass/Fail> screen.
Start the procedure from the <Target menu> - <Pass/Fail> screen.
: -0.2000 to 0.2000
: 0.00 to 20.00