… Calibration Interval Messages

If the instrument has not been used for an extended period of time since the previous measurement, a message
prompting white calibration will appear after start-up and before measurement. The time interval between the
previous calibration and when the message is displayed can be configured.
Memo
The interval until the calibration is displayed is set to "8 hours" when the instrument is shipped from the factory.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Calibration
interval", and then press the [Confirmation] key.
2
▲ and ▼ will be displayed above and below the
numbers representing the time until the message is
displayed. Use [▲] or [▼] to specify a value.
Settings
{ 1 to 24 h: Set the time interval between the previous
calibration and when the message should be
displayed to between 1 and 24 hours.
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Calibration setup> screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Calibration setup> screen.
Start the procedure from the <Calibration setup> screen.