Sets the threshold to be used for the sample SMC function.
Sets the number of measurements to perform when performing SMC average measurement. The average of the
data obtained from the measurements conducted by pressing the measurement button several times is
determined. The maximum number of measurements at this time is specified number in the setting + four times.
Memo
"3" is set as the number of measurements for SMC averaging when the instrument is shipped from the factory.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "SMC Times",
and then press the [Confirmation] key.
2
▲ and ▼ will be displayed above and below the
count in the count selection.
Use [▲] or [▼] to specify a value.
Settings
{ 3 to 10 times
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Measurement setup> screen.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Measurement setup> screen.
With SMC average measurement, measurement is performed for the configured number of times the
measurement button is pressed, and the average value of those measurements is saved as a single measurement.
After the set number of measurements has been performed, the average can be saved by selecting "Save".
To save the average automatically after the set number of measurements has been performed, select "Auto save"
under "Measurement cond." - "Measurement setup" - "SMC average option".