„ Pass/Fail
Edit the tolerance that will be used as the judgment criteria, and set the warning level and the parametric
coefficient default value.
If the color difference between the sample and the target exceeds the tolerance, the column of the display color
value relevant to the measurement display will be displayed in red. In addition, if even one display color value
exceeds the tolerance, judgment will be shown as "Fail".
If the color difference between the sample and the target exceeds the warning level amount, the column of the
display color value relevant to the measurement display will be displayed in yellow. In such cases, even if other
display color values do not exceed the tolerance, judgment will be shown as "Warning".
Memo • The default judgment criteria is set before judgment criteria for each target is set. For details, refer to P.73

"Default Data Settings".

• The tolerance setting items are the same as those for the color space or index currently selected for the observation
conditions.
• The optional Color Management Software "SpectraMagic NX" allows for easy configuration and use of judgment
criteria.
• Display the target to be set in advance before moving to <Target menu>.
Operating Procedure
1
Use [▲] or [▼] to move the cursor to "Pass/Fail",
and then press the [Confirmation] key to display the
<Pass/Fail> screen.
69
Start the procedure from the <Target menu> screen.