… Measurement Area Selection (CM-26dG/CM-26d)

Select the measurement area.
The selected measurement area can be checked on the status bar displayed on the screen. (Refer to P.23.)
Operating Procedure
1
Use the measurement area switch on the bottom of
the instrument to set the measurement area to MAV
or SAV.
Setting
{ MAV : Measurement area of 8 mm (10×7 mm with gloss measurement)
{ SAV
: Measurement area of 3 mm

… Target Mask Replacement (CM-26d/CM-25d)

Replace the target mask according to the sample to measure.
Notes
• Fit the target mask securely so that the
bottom is parallel to the instrument.
• Make sure the measurement area set for
the instrument matches the diameter of
the target mask being used.
Using the optional stapler type target mask (CM-
A268) makes it possible to accurately aim where
measurement will be performed.
33
Operate the measurement area switch on the instrument.
MAV
SAV