Download Print this page

Bard Composix L/P Instructions For Use Manual page 3

Low profi le polypropylene & eptfe mesh with positioning system
Hide thumbs Also See for Composix L/P:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
4) Directions for Using Composix™ L/P Mesh with Echo PS™ Positioning System
To activate and infl ate the device:
1.
Once the C
™ L/P Mesh with E
OMPOSIX
men, use a grasper to locate the blue retrieval loop on the infl ation tube, ensuring that the blue infl ation tube
is not wrapped around the mesh and is clearly visible (Fig. 9).
2.
Pass a suture passer device through healthy skin at the center of the hernia defect (avoid going directly
through the umbilicus). Grasp the blue retrieval loop and pull the retrieval loop and infl ation tube out of
the abdominal cavity (Fig. 10).
3.
Place an atraumatic clamp or hemostat on the infl ation tube at the level of the skin to temporarily hold the
device in place (Fig 11a). Cut the infl ation tube 1-2 cm below the retrieval loop with surgical scissors to
ensure the tube is unobstructed. Discard the retrieval loop (Fig. 11b).
4.
Remove the infl ation assembly from the sterile pouch and screw the infl ation adapter tightly to the syringe
(Fig. 12). Connect the infl ation tube and infl ation assembly as follows (Fig. 13):
a.
Ensure the clear cap of the adapter is pushed downward to open the infl ation tube channel
b.
Insert the infl ation tube at least 5 cm into the cap opening
c.
Push the clear cap of the adapter upward to lock in place
5.
To infl ate the device, release the clamp or hemostat and pull the infl ation tube upward to lift the mesh
off the viscera (Fig 14). Always grab and pull the tube directly. Do not lift up using the infl ation adapter/
assembly. Infl ate the device by pumping the syringe until the balloon is fully infl ated (Fig. 15). One to
three pumps will be required to fully infl ate the device depending on the size of the mesh. A slight high
pitch sound may occur; this is normal and indicates the infl ation assembly is working properly. If desired,
remove the syringe once infl ated by unscrewing from the adapter then set aside (Fig. 16).
Note: The mesh should now be unrolled, opened and ready to be properly positioned.
Fig. 9
Fig. 10
Fig. 13
Fig. 14
a
b
c
To properly position and fi xate the mesh:
6.
Raise the infl ation tube to properly adjust the mesh to the desired position then clamp the infl ation tube
to hold the device in place (Fig. 17).
7.
Use a grasper to orient the C
OMPOSIX
dark shaded areas on the device represent the long axis. The two shaded points at both ends of the long
axis indicate the midline of the C
OMPOSIX
8.
Once the device has been properly positioned and before the E
ensure that no tissue is entrapped between the device and abdominal wall and fi xate around the entire
patch perimeter of C
™ L/P Mesh with fasteners placed 1-2 cm apart and at least ½ cm inside
OMPOSIX
the outermost row of stitches (Fig. 24). Ensure that no fasteners are placed through the E
Positioning System (Fig. 19).
Fig. 17
Fig. 18
To defl ate and remove the E
PS™ Positioning System:
CHO
9.
To defl ate the E
PS™ Positioning System, release the clamp on the infl ation tube, cut the tube as
CHO
close to the skin as possible, and then discard (Fig. 20).
10. Begin removal of the E
PS™ Positioning System by grasping one of the two removal points marked by
CHO
the dark arrows adjacent to the B
®
ARD
motion (Fig. 21).
11. Continue to grasp the E
PS™ Positioning System, pulling it up to the tip of the trocar (Fig. 22). Re-
CHO
move both the E
PS™ Positioning System and trocar simultaneously (Fig. 23). Verify that the device
CHO
is fully intact after removal, and discard the E
12. Reinsert the trocar, place any additional fi xation as desired, and complete the procedure. Care should
be taken to ensure that the patch is adequately fi xated to the abdominal wall. If necessary, additional
fasteners and/or sutures should be placed.
Fig. 20
Fig. 21
PK3796862 BAW-35.indd 3
PK3796862 BAW-35.indd 3
PS™ Positioning System (the device) is inserted into the abdo-
CHO
Fig. 11
Fig. 12
b
a
Fig. 15
Fig. 16
™ L/P Mesh in relation to the defect (Fig. 18). The B
™ L/P Mesh.
PS™ Positioning System is defl ated,
CHO
Fig. 19
logo and pulling the positioning system off the mesh in one smooth
PS™ Positioning System appropriately.
CHO
Fig. 22
Fig. 23
Fixation
B
®
fi xation devices or nonabsorbable monofi lament sutures are recommended to properly secure
ARD
C
™ L/P Mesh. If other fi xation devices are used, they must be indicated for use in hernia repair. If
OMPOSIX
transfascial sutures are to be used for fi xation, place sutures after removing the E
from the body and after completing all mechanical fi xation. Care should be taken to ensure that the mesh is
adequately fi xated to the abdominal wall. If necessary, additional fasteners and/or sutures should be placed.
Reference Step #8 above.
Adverse Reactions
Possible complications include seroma, adhesions, hematomas, infl ammation, extrusion, fi stula formation and
recurrence of the hernia or soft tissue defect.
Traceability
Traceability labels which identify the type, size, and lot number of the device are attached to every package.
A label should be affi xed to the patient's permanent medical record to clearly identify the device which was
implanted.
If you experience a product failure, please contact Davol Inc. at 1-800-556-6275 for instructions on returning
the product.
Storage
C
™ L/P Mesh with E
OMPOSIX
CHO
room temperature until ready for use.
logo and
®
ARD
PS™
CHO
Bard, Composix, Davol, and Echo PS are trademarks and/or registered trademarks of C. R. Bard, Inc.
or an affi liate.
Copyright © 2010, 2011 C. R. Bard, Inc. All Rights Reserved.
CHO
Fig. 24
PS™ Positioning System should be stored in a clean, dry area at ambient
PS™ Positioning System
10/5/2011 1:01:27 PM
10/5/2011 1:01:27 PM

Advertisement

loading

This manual is also suitable for:

Echo ps