Table 11: Electromagnetic Immunity - Nellcor Oximax NPB-40 Operator's Manual

Handheld pulse oximeter
Hide thumbs Also See for Oximax NPB-40:
Table of Contents

Advertisement

Specifications

Table 11: Electromagnetic Immunity

The NPB-40 is intended for use in the electromagnetic environment
specified below. The customer or user of the NPB-40 should assure that
it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic
+ 6 kV contact
Discharge (ESD)
+ 8 kV air
IEC 61000-4-2
Electrical fast
+ 2 kV power
transient/burst
supply lines
IEC 61000-4-4
+ 1 kV for
input/output
lines
Power
3 A/m
Frequency
(50/60 Hz)
magnetic field
140
Compliance
Electromagnetic
Level
Environment
Guidance
Complies
Floors should be
wood, concrete, or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Complies
Main power should
be that of a typical
commercial or
hospital
environment.
Complies
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.