Initialization with Vial
Error ID: 34372
The autosampler attempted to initialize with a vial still in the gripper.
During initialization, the autosampler checks correct operation of the gripper
by closing and opening the gripper fingers while monitoring the motor
encoder. If a vial is still in the gripper when initialization is started, the
gripper fingers cannot close causing the error message to be generated.
Probable cause
1
Vial in Gripper
Error ID: 34308
The gripper arm attempted to move with a vial still in the gripper.
During specific stages of the sampling sequence, no vial should be held by the
gripper. The autosampler checks if a sample vial is stuck in the gripper by
closing and opening the gripper fingers while monitoring the motor encoder. If
the gripper fingers are unable to close, the error message is generated.
Probable cause
1
Agilent 1260 Infinity II Preparative Autosampler User Manual
Vial still in gripper.
Vial still in gripper.
Error Information
Sampler Error Messages
Suggested actions
Remove the vial using the Release Vial
function in the user interface. Reinitialize the
autosampler.
Suggested actions
Remove the vial using the Release Vial
function in the user interface. Reinitialize the
autosampler.