Maintenance
Exchange the Wash Port Assembly
9
Loosen the fixing screw.
11
Disconnect the wash tubing from the peristaltic pump.
13
Lift the flapper out of the bracket. Be careful not to
squeeze the wash tubing.
184
10
Lift out the needle.
12
Release the spring loaded pin.
14
Reconnect the wash tubing fitting to the new wash port
assembly.
Agilent 1260 Infinity II Preparative Autosampler User Manual