Leak Sensor Open

Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable cause
1
2
3
4
Agilent 1260 Infinity II Preparative Autosampler User Manual
Leak sensor not connected to the Power
Switch board.
Defective leak sensor.
Leak sensor incorrectly routed, being
pinched by a metal component.
Power switch assembly defective
Error Information
General Error Messages
Suggested actions
Ensure the leak sensor is connected correctly.
Exchange the leak sensor.
Exchange the leak sensor.
Exchange the power switch assembly.