Agilent Chemstation Setup - Agilent Technologies 1260 Infinity II User Manual

Infinitylab lc series preparative autosampler
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

13
LAN Configuration
PC and Agilent ChemStation Setup

Agilent ChemStation Setup

This procedure describes the Agilent ChemStation B.04.02 setup for the
1290 Infinity system using the 1290 Infinity DAD (G4212A) as the interfacing
module. The setup works in the same way for all other systems.
The LAN must be connected to detector due to high data load on communication to
N O T E
Control Software.
1 Open the ChemStation Configuration Editor.
2 Select from the menu Configure - Instruments.
3 Select Modular 3D LC System.
4 Give the Instrument a name.
Agilent 1260 Infinity II Preparative Autosampler User Manual