Maintenance
Exchange the Sample Loop Assembly
7
Reinstall the new sample loop.
N O T E
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
9
Turn or flap the wash port to the right.
11
Click Next until the needle assembly is moved to the
needle seat.
12
Select Back to finish the procedure.
14
Perform a pressure test.
166
8
In Lab Advisor use Next to lift the needle slowly into the
up position.
10
Install the safety cover (1) and fix the screw (2).
13
Close the doors.
Agilent 1260 Infinity II Preparative Autosampler User Manual
2
1