Needle Movement Failed; Missing Vial - Agilent Technologies 1260 Infinity II User Manual

Infinitylab lc series preparative autosampler
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Needle Movement Failed

Error ID: 34213, 34214
The needle arm failed to move down into the needle seat.
The lower position of the needle arm is monitored by a position sensor on the
sampling unit flex board. The sensor detects the successful completion of the
needle movement to the needle seat position. If the needle fails to reach the
end point, or if the sensor fails to recognize the needle arm movement, the
error message is generated.
Probable cause
1
2
3

Missing Vial

Error ID: 34305
No vial was found in the position defined in the method or sequence.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Probable cause
1
2
3
4
Agilent 1260 Infinity II Preparative Autosampler User Manual
Defective or dirty position sensor.
Defective motor.
Sticking spindle assembly.
No vial in the position defined in the
method or sequence.
Incorrect gripper alignment.
Defective gripper assembly (defective
gripper fingers or belt).
Defective transport assembly flex board.
Error Information
Sampler Error Messages
Suggested actions
Exchange the needle station board.
Exchange the needle drive motor.
Exchange the spindle assembly or the needle
station unit assembly.
Suggested actions
Install the sample vial in the correct position, or
edit the method or sequence accordingly.
Align gripper.
Exchange the gripper assembly.
Exchange the transport assembly.
8
137