Exchange the Gripper Arm

When
Defective gripper arm
Tools required
Description
Straightened paper clip.
Parts required
p/n
G1313-60010
Preparations
Finish any pending acquisition job
1
In the Local Controller start the maintenance mode and
select Change Gripper function.
OR
In the Tools section of the Agilent Lab Advisor software
select Service & Diagnostics > Maintenance Positions
> Change Gripper, click Start and wait until the gripper
is in maintenance position.
3
Identify the slit below the gripper motor and the gripper
arm release button.
Agilent 1260 Infinity II Preparative Autosampler User Manual
Description
Gripper assembly
2
Open the front door.
4
Rotate the arm approximately 2.5 cm (1 inch) to the left
and insert the straightened paper clip into the slit.
Maintenance

Exchange the Gripper Arm

9
177