Troubleshooting and Diagnostics
Maintenance Functions

Automatic Referencing

Automatic Referencing is required to compensate for small deviations in
positioning of the gripper which may occur after disassembling the module for
repair.
The Automatic Referencing procedure requires Calibration Tool
(G7129-60014) and uses several tray positions as reference points. Because
the tray is a rectangle, a four-point alignment is sufficient to correct all other
vial positions within the tray. On completion of the procedure, the corrected
gripper positions are stored in the instrument firmware.
Start
Moves the transport assembly and gripper arm to the position required to
install the calibration tools.
Next
Mount the calibration fixture. Position P1-C6 and P2-F6.
Please check help for further explanation.
N O T E
Next
Mount a 2 mL vial.
Next
Perform autoreferencing.
Next
Remove calibration fixture.
Back
Completes the procedure.