Agilent Technologies 1260 Infinity II User Manual page 146

Infinitylab lc series preparative autosampler
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

8
Error Information
Sample Cooler Error Messages
Cooler Condensate Error
Error ID: 30702, 30709, 30710
The sample cooler has detected condensate inside of the leak pane.
Probable cause
1
2
3
146
Blocked drainage
Defective condensate sensor
Defective leak sensor
Suggested actions
Switch off the instrument and remove the
condensate tubings. Place a beaker underneath
the waste outlet. Check the waste container.
There should be no kinked and sealed tubing.
For further information see
Cooler"
• Switch the sample cooler off/on and check
if after 8 min the error shows up again.
• Replace the condensate sensor.
Replace the sample cooler.
Agilent 1260 Infinity II Preparative Autosampler User Manual
"Install the Sample
on page 60.