Intermittent lock-ups with or without vial in the gripper fingers

With error messages
• motor overtemp (0 or 1 or 2 or 3)
• movement failed (0 or 1 or 2 or 3)
• missing vial
Personal damage, damage to the module
WA R N I N G
When a motor over temperature message has occurred, the sampler must be turned OFF
N O T E
for about 10 min to allow the motor to cool down.
1 Check the vials and the caps.
2 Reset the sampler and check the tension of the belts.
3 Exchange the gripper arm assembly (Gripper assembly (G1313-60010)), see
4 Start automatic referencing in Lab Advisor. For this procedure it is
5 Exchange Transport assembly ( Sample Transport Assembly
6 Exchange the Autosampler Main Board ( Autosampler Specific Board
7 For both steps (step 5 and 6), you have to do an automatic referencing again
Agilent 1260 Infinity II Preparative Autosampler User Manual
Troubleshooting Guide for the Sample Transport Assembly
Some of these procedures require a trained service engineer. Persons who are not
qualified MUST NOT perform these procedures.
For reliable operation, vials used with the autosampler must not have
tapered shoulders or caps that are wider than the body of the vial. For more
details, see
"List of Compatible Vials and Caps"
"Exchange the Gripper Arm"
mandatory to use the calibration tool.
On completion of the procedure, the corrected values are stored in the
instrument firmware.
(G7129-60600)).
(G7129-65800)) or the fusion board ( Fusion Board (G4200-65880)).
to store the correct values in the instrument firmware again.
Troubleshooting and Diagnostics
on page 48.
on page 177.