Exchange the Needle Seat Assembly
13
Connect the seat-capillary fitting to port 5 of the
injection valve.
15
Reconnect the sample-loop to the needle.
N O T E
Do not overtighten the fitting!
N O T E
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
162
14
Insert the needle (1). Align the needle in the seat, then
tighten the screw firmly (2).
16
In Lab Advisor use Next to lift the needle slowly into the
up position.
Agilent 1260 Infinity II Preparative Autosampler User Manual