5
Preparing the Module
Swage Fittings
5
Once you believe you have the fitting complete, loosen
the nut, and inspect the ferrule for the correct position
on the tubing.
Example of a perfect fitting
Examples of incorrect fittings
Figure 16
Examples of incorrect fittings
The first time that the swagelock fitting is used on a column or an injection valve, the
N O T E
position of the ferrule is permanently set. If changing from a column or an injection valve
to another, the fitting may leak or decrease the quality of the separation by contributing to
band broadening.
84
Agilent 1260 Infinity II Preparative Autosampler User Manual