1
In the Agilent Lab Advisor software select Service &
Diagnostics in the system screen (Tools) Maintenance
Positions > Change Needle/Loop, click Start and wait
until the needle assembly is in maintenance position.
3
Unscrew the safety cover screw.
Agilent 1260 Infinity II Preparative Autosampler User Manual
Exchange the Needle Assembly
2
Open the front door.
WA R N I N G
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
4
Remove the safety cover.
9
Maintenance
155