Exchange the Wash Port Assembly
19
In Lab Advisor use Next to lift the needle slowly into the
up position.
21
Install the safety cover (1) and fix the screw (2).
24
Close the doors.
186
2
1
Agilent 1260 Infinity II Preparative Autosampler User Manual
20
Turn or flap the wash port to the right.
22
Select Next to move the needle into the needle seat.
23
Select Back to finish the procedure.
25
Purge the wash port. Check if solvent will be delivered
from the peristaltic pump.