Intermittent lock-ups with or without vial in the gripper fingers

With error messages
• motor overtemp (0 or 1 or 2 or 3)
• movement failed (0 or 1 or 2 or 3)
• missing vial
Personal damage, damage to the module
WA R N I N G
When a motor over temperature message has occurred, the sampler must be turned OFF for
N O T E
about 10 minutes to allow the motor to cool down.
1 Check the vials and the caps.
2 Very heavy usage - use a macro.
3 Reset the sampler alignment to default value.
Agilent 1260 Infinity Standard Autosampler User Manual
Troubleshooting Guide for the Sample Transport Assembly
Some of these procedures require a trained service engineer. Persons who are not
qualified MUST NOT perform these procedures.
For reliable operation, vials used with the autosampler must not have
tapered shoulders or caps that are wider than the body of the vial. For more
details see the service note G1313-017.
A pre-sequence macro, inj_rset.mac will automatically reset the sampler at
the start of a sequence (ChemStation).
Reset tray alignment, and transport alignment is possible with the Control
Module and the ChemStation. To reset the transport alignment with the
ChemStation, enter following command in the command line.
Print sendmodule$(lals, "tray:alig 0.00,0.00")