Patient Data

Operating Status

60
Patient data is highlighted in the display above. Parameter values are
displayed directly beneath the heading for each parameter. In this
example, the %SpO
is 100, and the pulse rate is 190 beats per minute. An
2
asterisk indicates that the parameter is outside the set limits. If no data for
a parameter is available, three dashes (- - -) are displayed.
PA represents pulse amplitude. The number can range from 0 to 254.
There are no alarm parameters for this value. It can be used for trending
information as an indication of a change in pulse volume, relative pulse
strength, or circulation.
The Status column indicates alarm conditions and operating status of the
N-560. In this example, Pulse High (PH) means that the pulse rate upper
alarm limit has been exceeded. The status codes are listed in the following
table. As many as four codes can be displayed at one time in the Status
column.