10.1.3
When there is a Big Error
Flow chart 2
Big error
Display or print
value matching dot
prints value?
‚ x ‚ d ‚ r
Calibration
corrects?
‚ x ‚ d ‚ r
Scaling set
properly?
‚ x ‚ d ‚ r
Contact our dealer where you purchased
the instrument, or our sales representative.
10.1.4
When the Trend Recording Deflected
Flow chart 3
Trend Recording Deflected (0 or 100%)
Input signal wiring
correct?
‚ x ‚ d ‚ r
Value and sign
(+/-) of input
correct ?
‚ x ‚ d ‚ r
Range and scale
set properly
‚ x ‚ d ‚ r
Contact our dealer where you purchased
the instrument, or our sales representative.
(Dot point error)
‚ m ‚ n
‚ m ‚ n
Calibrate again
(See 9.5 on Page 91)
‚ x ‚ d ‚ r
‚ m ‚ n
Set properly
(See 7.1.1 on Page 46)
‚ m ‚ n
(Sensor disconnection, etc)
‚ m ‚ n
‚ m ‚ n
95• \
Calibrate
zero/span of chart
paper properly?
Wire correctly
(See 4.2
on Page 17 and 18)
Apply a correct input
(See 4.2
on Page 17 and 18)
Set correctly
(See 7.1.1 on Page 43)
‚ m ‚ n
Calibrate again
(See 7.2.11 on
Page 78)
HXPRM10mnCT001E