(1)• @ Calibration of Voltage
Conduct calibration in case of voltage, current or thermocouple input. For thermocouple input,
calibrate reference junction compensation together.
Wiring
Precision Voltage
(Current)
Generator
Conduct calibration according to the instructions in 7.2.12 Data Calibration (Calibration of the
voltage).(See page 79)
[Note]
‡ @ • @
Calibrate at any one channel.
‡ A • @ When a shunt resistor is attached to the input terminal block in case of current input, detach it
prior to calibration.
(2)• @ Calibration of Resistance temperature detector
Wiring
Precision Dial
Resistor
Conduct calibration according to the instructions in 7.2.12 Data Calibration (Calibration of the
resistance temperature detector).(See page 80)
[Note]
A resistance value should not disperse greater than 6 mƒ ¶
When calibrate RTD for a channel, the input terminals of the other channels are shortcut.
Recording requires RTD calibration for every channel
• ü
• ü
• ü
• ü
• {
• ü
• ü
• |
92• \
• {
Digital Voltmeter
• |
• {
TRM-10C Recorder
Input Terminals
• |
• ü
‚ `
TRM-10C
• ü
‚ a
Recorder
Input
Terminals
• ü
‚ a
from one wire to another.
HXPRM10mnCT001E