Replace The Nd Filter (Optional); Connect The Motorized Device Cable - Nikon Eclipse Ni-U Instruction Manual

Upright microscope assembly/maintenance
Table of Contents

Advertisement

19

Replace the ND filter (optional).

Tool: Hex driver (2 mm across flats)
The ND filter in the ND filter cassette of the Ni-U main
body can be replaced as follows:
(1)
Check that the microscope power switch is OFF.
(2)
Press all ND filter IN/OUT switches to the "IN" side.
(3)
Remove the rubber cap attached to the top of the
filter cassette cover (it can be removed with your
finger).
(4)
Insert a hex driver in the hole with the rubber cap
removed and loosen the internal screw. (The screw
is drop-proof and not detached from the cover.)
(5)
Hold the ND filter IN/OUT switch and pull out the
cassette.
(6)
Hold the end of the filter holding frame and widen
slightly to remove the filter.
(7)
When attaching the filter, place the filter on the
holding frame at the position indicated with the arrow
and slide it in.
20

Connect the motorized device cable.

Precautions for connecting cables
Make sure that the microscope and peripheral equipment are turned off before connecting
the cable.
When the microscope is motorized, see the figure on next page and connect the cable correctly.
• To connect the LAMP CTRL (BOX B) on the microscope to the LAMP CTRL (Ni-U) on the control box B, the lamp
control cable is required.
• To connect the DSC on the microscope to the DSC (Ni-U) on the control box B, the C-CTC camera trigger cable
L3/U3 is required.
• To connect the USB (H) on the DS-L3 to the USB on the control box B, a USB cable is required.
• To connect the EXT.I/O on the DS-L3/U3 to the DSC on the microscope or the Ex-OUT on the control box B,
the C-CTC camera trigger cable L3/U3 is required.
• To connect the motorized shutter to the EPI SHUTTER or DIA SHUTTER on the control box B, the NI-SHCL
motorized shutter cable long is required.
• To connect the RS232C on the C-HGFIE motorized HG precentered fiber illuminator to the HGFIE on the
control box B, an RS232C cross-wired cable is required (commercial, D-Sub9 pin female for both connectors).
Precautions for connecting the camera trigger cable
When connecting a camera trigger cable with the DSC connector, be sure to insert it completely.
Chapter 1
Assembly
(2)
Press all switches
to the "IN" side
22
(3)
(4)
(1)
OFF
Removing the filter cassette
(6)
(7)
Replacing the filter
Rubber cap