Maintenance &
6
Troubleshooting

Maintenance

The C-DiGit Blot Scanner requires only minimal maintenance. However, as with
any equipment utilizing electrical power, there is a danger of fire or electrical
shock if the equipment is not properly maintained.
Disconnect power before servicing. LI-COR Biosciences recommends that you
routinely inspect the system. The following are some general maintenance
guidelines:
Clean the exterior case parts with warm water and a damp cloth.
Clean the scanning surface with 100% methanol, then water.
CAUTION: DO NOT use acetone to clean the imaging surface.
Wipe all chemical spills from the case and scanning surface to prevent damage
to the surface coating.
Keep the scanning surface free of harsh solvents and and other combustibles.
Dry thoroughly before use.
Inspect all cables and power cords for evidence of fraying, exposed wire, or
loose connections.
Do not allow sample or sample covering to extend past the imaging surface.
This can prevent the instrument from operating properly by allowing light to
enter the scan area.
Troubleshooting – Hardware Problems

System Requirements

The C-DiGit Blot Scanner and Image Studio will run on Windows XP, Vista, and 7,
or on Mac OS 10.6, 10.7 and 10.8.

Cannot Connect to C-DiGit Blot Scanner

1. Install the Image Studio software before you try to connect the USB or Ethernet
cable to the C-DiGit Blot Scanner.
2. When using Windows XP, power on the C-DiGit Blot Scanner, allow the
drivers to install, and then reboot the computer. Wait for network connection,
and then connect to the C-DiGit Blot Scanner. See Appendix B for more
information.