Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual page 96

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

Chapter 3: Installing and Configuring the ITS
e) Select Upgrade if necessary from the APC drop-down menu to allow each
f) Select Smart Hopping-only Check and Update Configuration from the
g) Select Do Not Upgrade from the AP drop-down as you have only installed
4. Click Next> to continue.
The wizard automatically scans the IntelliVue Telemetry System for all
connected APCs. When the scan is completed (Figure 3-19), the following
message appears:
System [Philips] – Master APC [e.g., APC-02*] Complete
* Note that the displayed APC value will be different for each clinical site.
After scanning the ITS for connected APCs, the Upgrade Wizard will list any APCs
that have an incompatible configuration key. If such an APC is found, you cannot
continue the upgrade process. Locate the incompatible APC, disconnect it from
the ITS, reset it to its factory defaults, and then reconnect it to the ITS before
retrying the upgrade process.
3-18
Figure 3-6: APC/AP Firmware Selection Screen
APC to be upgraded. By default, an APC will not be upgraded if it is already
at the selected firmware version.
unlabeled drop-down menu to verify and update the APC configuration.
APCs on the ITS at this point.

Hide quick links: