Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual page 39

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

IntelliVue Telemetry System Infrastructure Installation and Service Guide
Table 1-13: UPS Specifications
UPS
Requirements
Input Voltage
120VAC +/-10%
Input
60Hz +/-3Hz minimum
Frequency
UPS AC
100, 120VAC model: one NEMA 5-15P plug.
Input Port
230 VAC model: one IEC 320 C14 plug.
Number of UPS
100, 120VAC model: 3 minimum (NEMA 5-15R output receptacles),
AC output ports
desirable to have 5 ports or greater.
230V model: 3 minimum (IEC 320 C13 output receptacles), desirable
to have 5 ports or greater.
Power delivery
UPS shall provide at least 225Watts for > 90 seconds to power
12 APs, 1 Sync Unit, 2 APC, 1 POE, & 1 Switch from one UPS unit with 5
UPS ports.
Or:
12 APs, 1 Sync Unit, 1 APC, 1 POE, from one UPS unit with 3 UPS ports.
Additional UPS would be needed to power redundant APC & Switch.
Power Fail
UPS must be responsive to power fail conditions such that ITS system
Transition
does not experience reboot due to transition to temporary battery
backup power.
Power Fail Alert
UPS shall provide local auditory response to power fail conditions.
EMI Filtering
Provide AC line EMI filtering over 100KHz to 30MHz on UPS ports.
EMC
FCC part 15 class A or B certification, CE Mark: CISPR 22/ EN 55022,
and CISPR 24 / EN 55024.
Safety
UL certification, EN 62040 (Low Voltage Directive).
Mounting
Rack Mounting is desirable.
IntelliVue Telemetry System Specifications
Parameter Specification
1.4 GHz ITS
100, 120, 127, 220, 240 VAC ± 10%
50, 60Hz +/-3Hz
2.4 GHz ITS
1-25

Hide quick links: