Figure 5: ITS PoE Unit to ITS Synchronization Unit Cable Connections

Step 3. Set Up Your Service PC.

To prepare your service PC to configure the APCs, you must:
• Configure your PC to connect to the ITS Wireless Subnet.
• Copy the Philips Upgrade Wizard tool and the latest ITS APC and AP firmware
files to your PC from the Philips Network Infrastructure Tools CD-ROM.

Configuring Your PC to Connect to the ITS Wireless Subnet

To configure your PC to connect to the ITS wireless subnet:
1. Configure your Service PC's network adapter for 100 Mbs full duplex
communications and assign a fixed IP address to your Service PC that falls
within the ITS subnet address space:
a) Disconnect all network connections from your PC.
b) Select Start/Settings/Network Connections/Local Area Connection from
c) Click Properties.
IntelliVue Telemetry System Infrastructure Installation and Service Guide