Note The APCs will appear in the APC web interface with the System ID of Philips. Do not
modify this setting.
5. Select the APC in the View Device tree (selected device is displayed green) and
then select the Configure tab.
In the AP Controller Network Configuration screen, configure the following
parameters as necessary:
• System ID - Verify setting is Philips. If not, change it to Philips.
• AP Controller Name - Enter APC Name from APC worksheet (page 2-29).
• IP Addressing - Verify that the Specify IP radio button is marked and that the
6. Click SAVE, and then verify the settings by selecting the Status tab.
7. At this time, add all other APCs to the network by repeating steps 1, 2, 3, 5, and
6 for each APC.
IntelliVue Telemetry System Infrastructure Installation and Service Guide
Newly Added
APC
Figure 3-4: APC Web Interface
IP Address, Subnet Mask and Default Gateway settings are as expected per
APC worksheet (page 2-29). If not, change the settings to match the
worksheet.