ITS Access Point Controller Test and Inspection Procedures

Table 5-3 lists which tests you must perform when performing service events on
IntelliVue Telemetry System Access Point Controllers.
Installation
Preventive Maintenance
Any component repair or replacement
Hardware Upgrade
Software Upgrade
All other Service Events
Table 5-4: ITS Access Point Controller Test and Inspection Matrix
Test Block
Name
APC Visual
APC Power On
APC Performance
APC
Revision Check
APC Safety
IntelliVue Telemetry System Infrastructure Installation and Service Guide
Table 5-3: ITS Access Point Controller Test and Inspection Requirements
When Performing this Service Event...
Test or Inspection to Perform
Inspect all system components for obvious
damage. Visually inspect all system
components, cables, and connectors. Check
for signs of abrasion, wear, or other damage.
With power connected to each active Network
device, observe that all lights visible on the
front panel are in proper status and that no
error conditions are shown. Following are
normal conditions for each type of device:
- Power LED - On GREEN
- Network Utilization LEDs -
See page 1-10.
- 100 BaseT LED - On GREEN
-LINK/Act - On primarily amber, flashes when
network activity is present
Does the APC appear in the APC web browser
management screens? Does data from a
wireless device associated with this APC
appear at the Information Center?
Does the revision reported in the APC web
browser screen match the revision loaded?
Run the Philips Upgrade Wizard to check and
verify final configurations.
No safety test is required
ITS Access Point Controller Test and Inspection Procedures
Complete These Required Test(s)
Visual, Power On, Performance
N/A
Power On, Performance
N/A
Power On, Performance, Revision Check
Visual, Power On, Performance
Expected Results
No visible damage
Devices power up into
expected status; no error
indications are shown.
Expected answers are
"Yes". If so, Performance
test passed.
Expected answer is
"Yes". If so, Revision
Check test passed.
Information to
Record on
Service Record
V:P or V:F
where:
P = Pass
F = Fail
PO:P or PO:F
where:
P = Pass
F = Fail
P:P or P:F
where:
P = Pass
F = Fail
RC:P or RC:F
where:
P = Pass
F = Fail
S:NA
where:
NA=Not
Applicable
5-3