Supported Topologies and System Limits

The IntelliVue Telemetry System is installed as part of an overall IntelliVue Clinical
Network installation. There are two basic IntelliVue Clinical Network topologies into
which the IntelliVue Telemetry System can be installed:
• non-routed ICN topology
• routed ICN topology

Installing the ITS within a Non-routed ICN

In a non-routed ICN configuration, the ICN functions as an independent network
with a single IntelliVue Database (DBS) Domain. Figure 1-10 represents an ITS
installed in a non-routed ICN.
Note the following guidelines for installing the ITS within a non-routed ICN topology:
• All ITS wireless devices (transceivers/wireless bedside monitors) must reside on
the ICN Database Domain on which the ITS infrastructure is installed.
• Up to 48 ITS Standard or Core Access Points may be installed on a non-routed
ICN topology.
• The 2.4 GHz ITS may only be installed on a non-routed ICN topology.