Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual page 60

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

Chapter 2: Planning Your ITS Deployment
-
-
For these reasons, care must be taken when defining the groups. APs which are not
physically located within a care area must still be given a "home" group for
equipment management purposes. This includes APs in transport areas as well as
AP in procedure areas (such as X-ray or Dialysis) that may be used by patients in
more than one care group, such as ICU and CCU.
For example, consider a hospital that has APs deployed in six main areas:
• ICU
• CCU
• X-Ray
• Dialysis
• Pulmonary
• Transport corridors in between these departments
In this hospital, the ICU and CCU are on two different ICNs, and there are no IICs in
any of the procedure areas (X-Ray, Pulmonary or Dialysis).
In planning the groups for this hospital, the service provider handling the
installation must decide where the alerts destination should be for each AP group
and also ensure that the group is also defined in Config Wizard and used similarly.
The service provider decides the groups should be set up as follows:
AP Location
ICU
CCU
X-Ray
Dialysis
Pulmonary
Transport
corridors
In this case, the ICU DBS will accumulate the alerts logs for all APs except those in
the CCU. All IICs in the ICU ICN will display alert messages when they occur (typically
only on device failure).
Conversely, only the CCU IIC will display messages for APs in the CCU and its DBS
will accumulate a database of alert data only for the CCU APs.
2-18
The DBS will accumulate all alerts for those APs and build an alert database
from them.
The DBS will send out alert system messages to all IICs on its ICN.
Table 2-6: Sample AP Groupings
Group Name
ICU
CCU
ICU
ICU
ICU
ICU
Config Wizard Devices
All beds, transmitters, central
stations and APs in ICU
All beds, transmitters, central
stations and APs in CCU
APs in X-Ray
APs in Dialysis
APs in Pulmonary
APs in transport corridors
System Devices in the
Philips IntelliVue
Telemetry System
APs in ICU
APs in CCU
APs in X-Ray
APs in Dialysis
APs in Pulmonary
APs in transport corridors

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents