Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual page 167

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

Consider an example (Figure A-5) where a hospital wants three different ITSs to
provide wireless coverage to three units located on different floors.
In this example, because the PCU and CCU are on adjacent floors, but the ER is
further away, the ITSs in PCU and CCU must be synchronized to each other, but the
ED system does not need to be synchronized to them.
Figure A-6: Sample Multiple ITS Sync Network and Equipment Label Requirements
Note that all three ITSs must have different RF Access Codes configured and unique
Equipment Labels assigned.
IntelliVue Telemetry System Infrastructure Installation and Service Guide
Figure A-5: Sample Multiple ITS Coverage Requirements
Slave Sync
Cable Delay Switch Set to 2
Sync Network Requirements for Multiple ITSs
Master Sync
Cable Delay Switch Set to 1
200 Ft. Sync Cable Length
Master Sync
Cable Delay Switch Set to 1
A-5