Chapter 3: Installing and Configuring the ITS
For Standard APs, be sure the AP's Ethernet DIP switch settings are set as
shown in Figure 3-2.
5. Connect the Uninterruptible Power Supply (UPS) to an AC power source.
6. Connect and power the ITS Power over Ethernet Unit.
a) For each 1.4/2.4 GHz Access Point you have installed, connect a Category 5
b) Connect the ITS PoE Unit power cord to an available power outlet on the
c) Verify that ITS Power over Ethernet (PoE) Unit powers up properly and that
1
2
1 2 3 4 5 6
Figure 4: ITS PoE Unit to ICN Network Switch Cable Connections
7. Connect and power the ITS Synchronization Unit.
a) Verify that for the Master ITS Synchronization Unit, the Master/Slave toggle
b) For each 1.4/2.4 Access Point you have installed, connect a Category 5 UTP
c) Connect the ITS Synchronization Unit power cord to an available power
d) Verify that the ITS Synchronization Unit powers up properly and that its
Figure 5 shows cable connections between the ITS Power over Ethernet Unit
and the ITS Synchronization Unit to support connection of two 1.4/2.4 GHz
Smart-hopping Access Points.
3-6
UTP patch cable between a Data Port on the PoE Unit and an available port
on the ICN Network Switch.
back of the UPS.
its power indicator lights green.
Figure 4 shows cable connections between the ITS Power over Ethernet
(PoE) Unit and the ICN Network Switch to support connection of two 1.4/2.4
GHz Smart-hopping Access Points.
ICN Network Switch
Data & Power Ports (To APs)
Data Ports (To ICN Network Switch)
ITS Power over Ethernet Unit
switch is set to Master.
All other Sync Units in the network should have the Master/Slave toggle
switch is set to Slave. All Slave Sync units should have their front-panel
Cable Delay Switch set as described in Table 2-5.
patch cable between a Data & Power Port on the PoE Unit and a From Hub
Port (i.e., top connectors) on the ITS Synchronization Unit.
outlet on the back of the UPS.
power indicator lights green.