Environmental Specifications - Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

Environmental Specifications

For EMC purposes component parts of the system conform to the requirements of
EN60601-1-2:2001.
In general, the ITS is designed for use in an indoor environment and operates over
an ambient temperature range of 0
transceiver.
The Access Point, Sync Unit, APC, and Power over Ethernet Unit, and switches are all
classified as ITE devices.
Operating Temperature
Operating Humidity
Storage Temperature
Storage Humidity
Altitude (Operating and
Non-operating)
IntelliVue Telemetry System Infrastructure Installation and Service Guide
Table 1-16: ITS Environmental Specifications
Environmental Test for ITS
IntelliVue Telemetry System Specifications
to 55
C, excluding the patient-worn
°
°
Test Type & Limits
°
°
0
to 55
C
°
°
< 95% RH at 40
C (104
F)/
non-condensing
°
°
-40
C to 60
C
°
< 90% RH at 60
C
3,048 meters (10,000 ft)
Applicable System
Element
Standard and Core
APs, Remote
Antennas, Sync
Unit, APC
Standard and Core
APs, Remote
Antennas, Sync
Unit, APC
Standard and Core
APs, Remote
Antennas, Sync
Unit, APC
Standard and Core
APs, Remote
Antennas, Sync
Unit, APC
Standard and Core
APs, Remote
Antennas, Sync
Unit, APC
1-27

Hide quick links: