Using The Upgrade Wizard - Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

Using the Upgrade Wizard

Caution Close the APC web interface before running the Philips Upgrade Wizard. Do not add
components to or remove components from the IntelliVue Telemetry System while
the wizard is running.
Follow these steps to upgrade ITS Access Point Controllers and Access Points using
the Philips Upgrade Wizard:
Philips Upgrade Wizard:
1. Run the Philips Upgrade Wizard by double-clicking the file:
C:\TelemetryUpgrader\upgrader.exe
The Upgrade Wizard splash screen (Figure 3-17) appears.
2. Click Next> to continue.
The APC/AP firmware selection screen (Figure 3-18) appears.
3. Complete the settings on the APC/AP firmware selection screen:
a) Click ... and then browse to the directory (e.g.,
b) Specify what type of APs you have installed by marking the appropriate radio
c) Verify that the correct firmware versions are displayed in the APC Version,
d) Select the ICN IP address (e.g., 172.31.240.4) you have configured for the
IntelliVue Telemetry System Infrastructure Installation and Service Guide
Figure 3-5: Upgrade Wizard Splash Screen
in which you have installed the APC and AP firmware
and APC Images\)
files.
button, 1.4 GHz Smart Hopping or 2.4 GHz Smart Hopping.
Additionally, if you have installed Core APs, then be sure to mark the System
includes Cluster APs (with or without Remote Antennas) checkbox.
AP Version (Standard), and AP Version (Cluster) fields.
support PC on which you are running the wizard from the Use IP Address
drop-down menu. The menu lists the IP addresses of all Ethernet NICs
configured on the PC.
Step 6. Run the Philips Upgrade Wizard.
C:\TelemetryUpgrader\AP
3-17