The GL450 device settings and email settings when an alarm has been generated are made here.
(1) Device Number
Specifies the Device Control Number for settings.
(2) Room Temp Compensation: Off, On
This function is only used when a temperature range has been specified. "On" is normally
selected. When making this setting, there is a choice between Internal (the GL450's room
temperature compensation settings are used) or External. Select External when you have the
necessary equipment and know-how.
(3) Burn Out: Off, On
This function is only used when a temperature range has been specified. It is normally set to
"On". When a thermocouple sensor has burnt out, that status is displayed on the screen.
(4) Output alarm when burn out (Alarm Number 4)
Alarm number 4 is output if a burnout occurred during temperature measurement.
(5) Temp Unit: Celsius (°C), Fahrenheit (°F)
Select the setting that you want to be displayed during temperature measurement.
(6) AC Line Cycley
Select the AC line frequency that matches the one normally used.
(7) Power On Start: Off, On
This function enables the settings that were in effect when measurement last ended to be used
to start measurement as soon as the power is switched on. (Data being measured using the
OPS022 software cannot be captured when the Power On Start function is used).
(8) Store the setting conditions into GL450: On (save settings), Off (do not save settings)
This parameter selects whether or not to save the software settings to the GL450.
(9) Synchronize PC and device clocks
Click this button to change the GL450's clock setting to match that of the computer.
(10) Factory Default Setting
This parameter returns all the settings to the factory defaults. Use this function when you want
to start from scratch.
4-28
CHECKPOINT
• If Power On Start is used to activate the GL450 and then the device is connected using the OPS022
application, the OPS022 status is displayed during data capture (waveforms are not displayed).
• If Power On Start is enabled while the device is being connected using the OPS022 application,
restart the OPS022 connection procedure (disconnect and then reconnect).